Change calibration algorithm #22
@ -132,7 +132,7 @@ and
|
|||||||
.. doxygenfunction:: adc_pt1000_get_resistance_calibration
|
.. doxygenfunction:: adc_pt1000_get_resistance_calibration
|
||||||
:outline:
|
:outline:
|
||||||
|
|
||||||
are used to set the reistance calibration internally. For a guide on how to calibrate the deivce, see the corresponding :ref:`usage_calibration` usage page.
|
are used to set the resistance calibration internally. For a guide on how to calibrate the deivce, see the corresponding :ref:`usage_calibration` usage page.
|
||||||
|
|
||||||
The calibration is calculated the following way:
|
The calibration is calculated the following way:
|
||||||
|
|
||||||
@ -147,13 +147,13 @@ The calibration is calculated the following way:
|
|||||||
OFFSET [label="Offset", description="Offset Correction :math:`O`"];
|
OFFSET [label="Offset", description="Offset Correction :math:`O`"];
|
||||||
OUT [shape=endpoint, description="Corrected Value"];
|
OUT [shape=endpoint, description="Corrected Value"];
|
||||||
|
|
||||||
LF -> SENS -> OFFSET -> OUT
|
LF -> OFFSET -> SENS -> OUT
|
||||||
}
|
}
|
||||||
|
|
||||||
The final calibrated PT1000 resistance is calculated as:
|
The final calibrated PT1000 resistance is calculated as:
|
||||||
|
|
||||||
.. math::
|
.. math::
|
||||||
R_{PT1000_{corr}} = R_{PT1000_{LF}} \cdot (1 + \sigma) + O
|
R_{PT1000_{corr}} = (R_{PT1000_{LF}} - O) \cdot (1 + \sigma)
|
||||||
|
|
||||||
The default values, if no calibration is loaded / executed, are:
|
The default values, if no calibration is loaded / executed, are:
|
||||||
|
|
||||||
|
Loading…
Reference in New Issue
Block a user